|
| Characteristics |
| Principle
and accuracy of the polarimeter |
|
The
microPOL polarimeter operates in five channels centered at
0.49, 0.67, 0.865, 1.6 and 2.2 µm. This instrument was
designed, initially, to study mainly the polarization properties
of land surfaces from the visible to middle infrared spectral
range.

Figure 1 : View of the microPOL system (source. F.
Auriol)
It is composed of 15 separate collimators
associated to 15 distinct (lens-filter-polarizer-detector)
blocks (fig. 1). The principle of polarization measurement
is the same as for POLDER-1/2 (fig. 2).

Figure 2 : Schematic principle of polarization measurements
Three polarizer orientations (shifted by
60°) at each wavelength give three independant measurements
which then are combined to derive total radiance L, polarized
radiance Lp as well as polarisation angle for a given viewing
geometry.
The absolute calibration accuracy is about 3% for radiance
and polarized radiance at wavelength smaller than 0.9 µm
and about 5-7 % for the longer wavelengths. In the 1.6 and
2.2 µm channels detectors are stabilized at a temperature
of -10°C (InGaAs detectors). It has been shown that
the effect of temperature variation on calibration is smaller
than 2% at maximum. This instrument has a narrow IFOV about
of 2° (fig. 3).

Figure 3 : MicroPOL fied of view
When this instrument
is setup in aircraft, we also have the possibility
to select several viewing angles (0°, 15°,
30° and 45°) which allows the selection of
the more pertinent scattering angle with respect to
the experiment objectives. Acquisition time for one
detector is 0.5 ms, so that the set of total and polarized
radiances at the 5 wavelengths needs no more than
7.5 ms. The radiometric accuracy (NeΔL) is 2
10-5. |
|
|
|
|
|
| |